Unknown

Dataset Information

Susceptibility of HPV16 and 18 to high level disinfectants indicated for semi-critical ultrasound probes.


ABSTRACT: Ultrasound probes used in endocavitary procedures have been shown to be contaminated with high-risk HPV after routine use and HPV is also known to be resistant to some high level disinfectants (HLDs). This study compared efficacy of two leading ultrasound probe HLD methods; liquid ortho-phthalaldehyde (Cidex® OPA) and an automated device using sonicated hydrogen peroxide (trophon® EPR) against HPV16 and HPV18 in a hard-surface carrier test. Native HPV16 and HPV18 virions were generated in organotypic epithelial raft cultures. Viral lysates were dried onto carriers with a 5% (v/v) protein soil. Efficacy tests were performed against the automated device at 35% and 31.5% H2 O2 and 0.55% OPA in quadruplicate with matched input, neutralization, and cytotoxicity controls. Hypochlorite was includ

SUBMITTER: Ryndock E 

PROVIDER: S-EPMC5063110 | biostudies-literature | 2016 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Sorry, this publication's infomation has not been loaded in the Indexer, please go directly to PUBMED or Altmetric.

Similar Datasets