Optimal configuration of respiratory navigator gating for the quantification of left ventricular strain using spiral cine displacement encoding with stimulated echoes (DENSE) MRI.


ABSTRACT:

Purpose

To determine the optimal respiratory navigator gating configuration for the quantification of left ventricular strain using spiral cine displacement encoding with stimulated echoes (DENSE) MRI.

Materials and methods

Two-dimensional spiral cine DENSE was performed on a 3 Tesla MRI using two single-navigator configurations (retrospective, prospective) and a combined "dual-navigator" configuration in 10 healthy adults and 20 healthy children. The adults also underwent breathhold DENSE as a reference standard for comparisons. Peak left ventricular strains, signal-to-noise ratio (SNR), and navigator efficiency were compared. Subjects also underwent dual-navigator gating with and without visual feedback to determine the effect on navigator efficiency.

Results

There were no differences in circumferential, radial, and longitudinal strains between navigator-gated and breathhold DENSE (P = 0.09-0.95) (as confidence intervals, retrospective: [-1.0%-1.1%], [-7.4%-2.0%], [-1.0%-1.2%]; prospective: [-0.6%-2.7%], [-2.8%-8.3%], [-0.3%-2.9%]; dual: [-1.6%-0.5%], [-8.3%-3.2%], [-0.8%-1.9%], respectively). The dual configuration maintained SNR compared with breathhold acquisitions (16 versus 18, P = 0.06). SNR for the prospective configuration was lower than for the dual navigator in adults (P = 0.004) and children (P < 0.001). Navigator efficiency was higher (P < 0.001) for both retrospective (54%) and prospective (56%) configurations compared with the dual configuration (35%). Visual feedback improved the dual configuration navigator efficiency to 55% (P < 0.001).

Conclusion

When quantifying left ventricular strains using spiral cine DENSE MRI, a dual navigator configuration results in the highest SNR in adults and children. In adults, a retrospective configuration has good navigator efficiency without a substantial drop in SNR. Prospective gating should be avoided because it has the lowest SNR. Visual feedback represents an effective option to maintain navigator efficiency while using a dual navigator configuration.

Level of evidence

2 J. Magn. Reson. Imaging 2017;45:786-794.

SUBMITTER: Hamlet SM 

PROVIDER: S-EPMC5269554 | biostudies-literature | 2017 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Optimal configuration of respiratory navigator gating for the quantification of left ventricular strain using spiral cine displacement encoding with stimulated echoes (DENSE) MRI.

Hamlet Sean M SM   Haggerty Christopher M CM   Suever Jonathan D JD   Wehner Gregory J GJ   Andres Kristin N KN   Powell David K DK   Zhong Xiaodong X   Fornwalt Brandon K BK  

Journal of magnetic resonance imaging : JMRI 20160726 3


<h4>Purpose</h4>To determine the optimal respiratory navigator gating configuration for the quantification of left ventricular strain using spiral cine displacement encoding with stimulated echoes (DENSE) MRI.<h4>Materials and methods</h4>Two-dimensional spiral cine DENSE was performed on a 3 Tesla MRI using two single-navigator configurations (retrospective, prospective) and a combined "dual-navigator" configuration in 10 healthy adults and 20 healthy children. The adults also underwent breathh  ...[more]

Similar Datasets

| S-EPMC5125889 | biostudies-literature
| S-EPMC2946451 | biostudies-literature
| S-EPMC3480812 | biostudies-literature
| S-EPMC7897469 | biostudies-literature
| S-EPMC8295221 | biostudies-literature
| S-EPMC5028227 | biostudies-literature
| S-EPMC6107388 | biostudies-literature
| S-EPMC6816028 | biostudies-literature
| S-EPMC7580660 | biostudies-literature
| S-EPMC6525145 | biostudies-literature