Evaluating the accuracy of geometrical distortion correction of magnetic resonance images for use in intracranial brain tumor radiotherapy.


ABSTRACT:

Aim

Determine the 1) effectiveness of correction for gradient-non-linearity and susceptibility effects on both QUASAR GRID3D and CIRS phantoms; and 2) the magnitude and location of regions of residual distortion before and after correction.

Background

Using magnetic resonance imaging (MRI) as a primary dataset for radiotherapy planning requires correction for geometrical distortion and non-uniform intensity.

Materials and methods

Phantom Study: MRI, computed tomography (CT) and cone beam CT images of QUASAR GRID3D and CIRS head phantoms were acquired. Patient Study: Ten patients were MRI-scanned for stereotactic radiosurgery treatment. Correction algorithm: Two magnitude and one phase difference image were acquired to create