Intrafractional 6D head movement increases with time of mask fixation during stereotactic intracranial RT-sessions.


ABSTRACT:

Background

The present study investigates the intrafractional accuracy of a frameless thermoplastic mask used for head immobilization during stereotactic radiotherapy. Non-invasive masks cannot completely prohibit head movements. Previous studies attempted to estimate the magnitude of intrafractional inaccuracy by means of pre- and postfractional measurements only. However, this might not be sufficient to accurately map also intrafractional head movements.

Materials and methods

Intrafractional deviation of mask-fixed head positions was measured in five patients during a total of 94 fractions by means of close-meshed repeated ExacTrac measurements (every 1.4 min) conducted during the entire treatment session. A median of six (range: 4 to 11) measurements were recorded per fraction, delivering a dataset of 453 measurements.

Results

Random errors (SD) for the x, y and z axes were 0.27 mm, 0.29 mm and 0.29 mm, respectively. Median 3D deviation was 0.29 mm. Of all 3D intrafractional motions, 5.5 and 0.4% exceeded 1 mm and 2 mm, respectively. A moderate correlation between treatment duration and mean 3D displacement was determined (rs = 0.45). Mean 3D deviation increased from 0.21 mm (SD = 0.26 mm) in the first 2 min to a maximum of 0.53 mm (SD = 0.31 mm) after 10 min of treatment time.

Conclusion

Pre- and post-treatment measurement is not sufficient to adequately determine the range of intrafractional head motion. Thermoplastic masks provide both reliable interfractional and intrafractional immobilization for image-guided stereotactic hypofractionated radiotherapy. Greater positioning accuracy may be obtained by reducing treatment duration (< 6 min) and applying intrafractional correction.

Trial registration

Clinicaltrials.gov, NCT03896555, Registered 01 April 2019 - retrospectively registered.

SUBMITTER: Mangesius J 

PROVIDER: S-EPMC6921566 | biostudies-literature | 2019 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Intrafractional 6D head movement increases with time of mask fixation during stereotactic intracranial RT-sessions.

Mangesius Julian J   Seppi Thomas T   Weigel Rocco R   Arnold Christoph Reinhold CR   Vasiljevic Danijela D   Goebel Georg G   Lukas Peter P   Ganswindt Ute U   Nevinny-Stickel Meinhard M  

Radiation oncology (London, England) 20191218 1


<h4>Background</h4>The present study investigates the intrafractional accuracy of a frameless thermoplastic mask used for head immobilization during stereotactic radiotherapy. Non-invasive masks cannot completely prohibit head movements. Previous studies attempted to estimate the magnitude of intrafractional inaccuracy by means of pre- and postfractional measurements only. However, this might not be sufficient to accurately map also intrafractional head movements.<h4>Materials and methods</h4>In  ...[more]

Similar Datasets

| S-EPMC11524472 | biostudies-literature
| S-EPMC9361321 | biostudies-literature
| S-EPMC4775825 | biostudies-other
| S-EPMC7716004 | biostudies-literature
| S-EPMC7807636 | biostudies-literature
| S-EPMC9709257 | biostudies-literature
| S-EPMC5777665 | biostudies-literature
| S-EPMC11899069 | biostudies-literature
| PRJEB45035 | ENA
| S-EPMC6753355 | biostudies-literature