ABSTRACT: Objectives
To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio™ COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients (n = 412) attending primary healthcare centres.Methods
RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimen collection. Samples displaying discordant results were processed for culture in Vero E6 cells. Presence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in cell cultures was confirmed by RT-PCR.Results
Out of 412 patients, 43 (10.4%) tested positive by RT-PCR and RAD, and 358 (86.9%) tested negative by both methods; discordant results (RT-PCR+/RAD-) were obtained in 11 patients (2.7%). Overall specificity and sensitivity of rapid antigen detection (RAD) was 100% (95%CI 98.7-100%) and 79.6% (95%CI 67.0-88.8%), respectively, taking RT-PCR as the reference. Overall RAD negative predictive value for an estimated prevalence of 5% and 10% was 99% (95%CI 97.4-99.6%) and 97.9% (95%CI 95.9-98.9), respectively. SARS-CoV-2 could not be cultured from specimens yielding RT-PCR+/RAD- results (n = 11).Conclusion
The Panbio™ COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres. More crucially, the data suggested that patients with RT-PCR-proven COVID-19 testing negative by RAD are unlikely to be infectious.
SUBMITTER: Albert E
PROVIDER: S-EPMC7662075 | biostudies-literature | 2021 Mar
REPOSITORIES: biostudies-literature
Albert Eliseo E Torres Ignacio I Bueno Felipe F Huntley Dixie D Molla Estefanía E Fernández-Fuentes Miguel Ángel MÁ Martínez Mireia M Poujois Sandrine S Forqué Lorena L Valdivia Arantxa A Solano de la Asunción Carlos C Ferrer Josep J Colomina Javier J Navarro David D
Clinical microbiology and infection : the official publication of the European Society of Clinical Microbiology and Infectious Diseases 20201113 3
<h4>Objectives</h4>To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio™ COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients (n = 412) attending primary healthcare centres.<h4>Methods</h4>RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out wit ...[more]