One-year MR angiographic and clinical follow-up after intracranial mechanical thrombectomy using a stent retriever device.


ABSTRACT:

Background and purpose

Little is known about the consequences of arterial wall damage that may be due to mechanical endovascular thrombectomy. Our aim was to perform 1-year MR angiographic and clinical follow-up of patients treated with mechanical endovascular thrombectomy using the Solitaire device.

Materials and methods

Patients with stroke treated between August 2010 and July 2012 were prospectively evaluated with a minimum follow-up of 1 year after mechanical endovascular thrombectomy. Angiographic follow-up was performed on a 3T MR imaging scanner and included intracranial artery TOF MRA and supra-aortic artery gadolinium-enhanced MRA. Images were assessed to detect arterial abnormalities (stenosis, occlusion, dilation) and were compared with the final post-mechanical endovascular thrombectomy run to differentiate delayed and pre-existing abnormalities. Clinical evaluation was performed with the mRS and the 36-Item Short-Form Health Survey questionnaire quality-of-life scale.

Results

Thirty-nine patients were angiographically assessed at the mean term of 19 ± 4 months. MRA showed intracranial artery abnormalities in 10 patients, including 5 delayed intracranial artery abnormalities in 4 patients (4 stenoses and 1 dilation), 4 cases of pre-existing intracranial artery stenosis, and 2 occlusions. Pre-existing etiologic cervical artery stenosis or occlusion was observed in 2 patients. All these patients remained asymptomatic during the follow-up period. A significant clinical improvement was observed at 1-year follow-up in comparison with 3-month follow-up (P < .0001), with a good outcome achieved in 62.5% of patients and an acceptable quality of life restored.

Conclusions

One-year follow-up identifies delayed asymptomatic arterial abnormalities in patients treated with the Solitaire device.

SUBMITTER: Eugene F 

PROVIDER: S-EPMC7965909 | biostudies-literature | 2015 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

One-year MR angiographic and clinical follow-up after intracranial mechanical thrombectomy using a stent retriever device.

Eugène F F   Gauvrit J-Y JY   Ferré J-C JC   Gentric J-C JC   Besseghir A A   Ronzière T T   Raoult H H  

AJNR. American journal of neuroradiology 20140814 1


<h4>Background and purpose</h4>Little is known about the consequences of arterial wall damage that may be due to mechanical endovascular thrombectomy. Our aim was to perform 1-year MR angiographic and clinical follow-up of patients treated with mechanical endovascular thrombectomy using the Solitaire device.<h4>Materials and methods</h4>Patients with stroke treated between August 2010 and July 2012 were prospectively evaluated with a minimum follow-up of 1 year after mechanical endovascular thro  ...[more]

Similar Datasets

| S-EPMC5136713 | biostudies-literature
| S-EPMC8115375 | biostudies-literature
| S-EPMC7368707 | biostudies-literature
| S-EPMC7771483 | biostudies-literature
| S-EPMC5698357 | biostudies-literature
| S-EPMC9898825 | biostudies-literature
| S-EPMC3465948 | biostudies-literature
| S-EPMC7655352 | biostudies-literature
| S-EPMC8421856 | biostudies-literature
| S-EPMC10800261 | biostudies-literature