Performance of antigen testing for diagnosis of COVID-19: a direct comparison of a lateral flow device to nucleic acid amplification based tests.


ABSTRACT:

Objectives

The gold standard for diagnosing an infection with SARS-CoV-2 is detection of viral RNA by nucleic acid amplification techniques. Test capacities, however, are limited. Therefore, numerous easy-to-use rapid antigen tests based on lateral flow technology have been developed. Manufacturer-reported performance data seem convincing, but real-world data are missing.

Methods

We retrospectively analysed all prospectively collected antigen tests results performed between 23.06.2020 and 26.11.2020, generated by non-laboratory personnel at the point-of-care from oro- or nasopharyngeal swab samples at the University Hospital Augsburg and compared them to concomitantly (within 24 h.) generated results from molecular tests.

Results

For a total of 3630 antigen tests, 3110 NAAT results were available. Overall, sensitivity, specificity, NPV and PPV of antigen testing were 59.4%, 99.0%, 98.7% and 64.8%, respectively. Sensitivity and PPV were lower in asymptomatic patients (47.6% and 44.4%, respectively) and only slightly higher in patients with clinical symptoms (66.7% and 85.0%, respectively). Some samples with very low Ct-values (minimum Ct 13) were not detected by antigen testing. 31 false positive results occurred. ROC curve analysis showed that reducing the COI cut-off from 1, as suggested by the manufacturer, to 0.9 is optimal, albeit with an AUC of only 0.66.

Conclusion

In real life, performance of lateral-flow-based antigen tests are well below the manufacturer's specifications, irrespective of patient's symptoms. Their use for detection of individual patients infected with SARS-CoV2 should be discouraged. This does not preclude their usefulness in large-scale screening programs to reduce transmission events on a population-wide scale.

SUBMITTER: Kahn M 

PROVIDER: S-EPMC8354301 | biostudies-literature | 2021 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of antigen testing for diagnosis of COVID-19: a direct comparison of a lateral flow device to nucleic acid amplification based tests.

Kahn Maria M   Schuierer Lukas L   Bartenschlager Christina C   Zellmer Stephan S   Frey Ramona R   Freitag Marie M   Dhillon Christine C   Heier Margit M   Ebigbo Alanna A   Denzel Christian C   Temizel Selin S   Messmann Helmut H   Wehler Markus M   Hoffmann Reinhard R   Kling Elisabeth E   Römmele Christoph C  

BMC infectious diseases 20210810 1


<h4>Objectives</h4>The gold standard for diagnosing an infection with SARS-CoV-2 is detection of viral RNA by nucleic acid amplification techniques. Test capacities, however, are limited. Therefore, numerous easy-to-use rapid antigen tests based on lateral flow technology have been developed. Manufacturer-reported performance data seem convincing, but real-world data are missing.<h4>Methods</h4>We retrospectively analysed all prospectively collected antigen tests results performed between 23.06.  ...[more]

Similar Datasets

| S-EPMC10077809 | biostudies-literature
| S-EPMC9022615 | biostudies-literature
| S-EPMC9283405 | biostudies-literature
| S-EPMC9443926 | biostudies-literature
| S-EPMC4109970 | biostudies-literature
| S-EPMC10858045 | biostudies-literature
| S-EPMC5280847 | biostudies-literature
| S-EPMC5648715 | biostudies-literature
| S-EPMC11674760 | biostudies-literature
| S-EPMC8864475 | biostudies-literature