Unknown

Dataset Information

0

Leadless left ventricular endocardial pacing for CRT upgrades in previously failed and high-risk patients in comparison with coronary sinus CRT upgrades.


ABSTRACT:

Aims

Cardiac resynchronization therapy (CRT) upgrades may be less likely to improve following intervention. Leadless left ventricular (LV) endocardial pacing has been used for patients with previously failed CRT or high-risk upgrades. We compared procedural and long-term outcomes in patients undergoing coronary sinus (CS) CRT upgrades with high-risk and previously failed CRT upgrades undergoing LV endocardial upgrades.

Method and results

Prospective consecutive CS upgrades between 2015 and 2019 were compared with those undergoing WiSE-CRT implantation. Cardiac resynchronization therapy response at 6 months was defined as improvement in clinical composite score (CCS) and a reduction in LV end-systolic volume (LVESV) ≥15%. A total of 225 patients were analysed; 121 CS and 104 endocardial upgrades. Patients receiving WiSE-CRT tended to have more comorbidities and were more likely to have previous cardiac surgery (30.9% vs. 16.5%; P = 0.012), hypertension (59.2% vs. 34.7%; P < 0.001), chronic obstructive airways disease (19.4% vs. 9.9%; P = 0.046), and chronic kidney disease (46.4% vs. 21.5%; P < 0.01) but similar LV ejection fraction (30.0 ± 8.3% vs. 29.5 ± 8.6%; P = 0.678). WiSE-CRT upgrades were successful in 97.1% with procedure-related mortality in 1.9%. Coronary sinus upgrades were successful in 97.5% of cases with a 2.5% rate of CS dissection and 5.6% lead malfunction/displacement. At 6 months, 91 WiSE-CRT upgrades and 107 CS upgrades had similar improvements in CCS (76.3% vs. 68.5%; P = 0.210) and reduction in LVESV ≥15% (54.2% vs. 56.3%; P = 0.835).

Conclusion

Despite prior failed upgrades and high-risk patients with more comorbidities, WiSE-CRT upgrades had high rates of procedural success and similar improvements in CCS and LV remodelling with CS upgrades.

SUBMITTER: Sidhu BS 

PROVIDER: S-EPMC8502498 | biostudies-literature | 2021 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Leadless left ventricular endocardial pacing for CRT upgrades in previously failed and high-risk patients in comparison with coronary sinus CRT upgrades.

Sidhu Baldeep Singh BS   Sieniewicz Benjamin B   Gould Justin J   Elliott Mark K MK   Mehta Vishal S VS   Betts Timothy R TR   James Simon S   Turley Andrew J AJ   Butter Christian C   Seifert Martin M   Boersma Lucas V A LVA   Riahi Sam S   Neuzil Petr P   Biffi Mauro M   Diemberger Igor I   Vergara Pasquale P   Arnold Martin M   Keane David T DT   Defaye Pascal P   Deharo Jean-Claude JC   Chow Anthony A   Schilling Richard R   Behar Jonathan M JM   Leclercq Christophe C   Auricchio Angelo A   Niederer Steven A SA   Rinaldi Christopher A CA  

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 20211001 10


<h4>Aims</h4>Cardiac resynchronization therapy (CRT) upgrades may be less likely to improve following intervention. Leadless left ventricular (LV) endocardial pacing has been used for patients with previously failed CRT or high-risk upgrades. We compared procedural and long-term outcomes in patients undergoing coronary sinus (CS) CRT upgrades with high-risk and previously failed CRT upgrades undergoing LV endocardial upgrades.<h4>Method and results</h4>Prospective consecutive CS upgrades between  ...[more]

Similar Datasets

| S-EPMC7397503 | biostudies-literature
| S-EPMC9794756 | biostudies-literature
| S-EPMC10616611 | biostudies-literature
| S-EPMC7397521 | biostudies-literature
| S-EPMC8647921 | biostudies-literature
| S-EPMC4913734 | biostudies-literature
| S-EPMC11385406 | biostudies-literature
| S-EPMC6275469 | biostudies-literature