Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures.


ABSTRACT:

Background

Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement.

Methods

We developed an investigational interface device to connect needle-electrodes to BIS sensors. The safety and clinical performance were investigated in patients who underwent surgery. Direct BIS values from a disposable BIS electrode and indirect values via the interface device were simultaneously recorded from the same areas of electrode placement in a single patient. The agreement between the direct and indirect BIS values was statistically analyzed.

Results

The interface device with a silver electrode demonstrated sufficient electric conduction to transmit electroencephalogram signals. The overall BIS curves were similar to those of direct BIS monitoring. Direct and indirect BIS values from 18 patients were statistically analyzed using a linear mixed model and a significant concordance was confirmed (indirect BIS = 7.0405 + 0.8286 * direct BIS, p<0.0001). Most observed data (2582/2787 data points, 92.64%) had BIS unit differences of 10 or less.

Conclusions

The interface device provides an opportunity for intraoperative BIS monitoring of patients, whose clinical situation does not permit the placement of conventional adhesive sensors at the standard location.

SUBMITTER: Harada H 

PROVIDER: S-EPMC8530286 | biostudies-literature | 2021

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of BIS monitoring using a novel interface device connecting conventional needle-electrodes and BIS sensors during frontal neurosurgical procedures.

Harada Hideki H   Muta Seiya S   Kakuma Tatsuyuki T   Ukeda Misa M   Ota So S   Hirata Maiko M   Fujioka Hiroshi H   Nakashima Osamu O   Dietel Barbara B   Tauchi Miyuki M  

PloS one 20211021 10


<h4>Background</h4>Bispectral index (BIS) monitoring is a widely used non-invasive method to monitor the depth of anesthesia. However, in the event of surgeries requiring a frontal approach, placement of the electrode may be impossible at the designated area to achieve a proper BIS measurement.<h4>Methods</h4>We developed an investigational interface device to connect needle-electrodes to BIS sensors. The safety and clinical performance were investigated in patients who underwent surgery. Direct  ...[more]

Similar Datasets

| S-EPMC8455790 | biostudies-literature
| S-EPMC5508692 | biostudies-literature
| S-EPMC11611908 | biostudies-literature
| S-EPMC7966422 | biostudies-literature
| S-EPMC7959833 | biostudies-literature
| S-EPMC4318734 | biostudies-other
| S-EPMC5997185 | biostudies-literature
| S-EPMC3638030 | biostudies-literature
| S-EPMC6191486 | biostudies-literature
| S-EPMC9879628 | biostudies-literature