Experimental ex-vivo performance study comparing a novel, pulsed thulium solid-state laser, chopped thulium fibre laser, low and high-power holmium:YAG laser for endoscopic enucleation of the prostate.


ABSTRACT:

Purpose

The aim of this study was to compare the enucleation performances of four different types of laser devices in an ex-vivo experiment: a novel, pulsed Tm:YAG solid-state laser evaluation model (p-Tm:YAG), chopped thulium fibre laser (TFL), low-power Ho:YAG laser (LP-Ho:YAG), and a high-power Ho:YAG laser (HP-Ho:YAG).

Methods

Our primary aim was to endoscopically separate the fascial layers of a porcine belly using laser fibres within a time period of 60 s. The size of a "tissue pocket" was assessed numerically. The enucleation characteristics reflecting the surgeon's experience were evaluated via the NASA Task Load Index (TLX) questionnaire and a questionnaire based on Likert scale.

Results

HP-Ho:YAG achieved with the available laser settings the largest overall