Early results after aortic annuloplasty with a complete external Dacron band.


ABSTRACT:

Objective

This study evaluates the early results of our initial experience with aortic annuloplasty using a complete external Dacron band in the setting of type Ic or type II aortic regurgitation (AR).

Methods

From May 2017 to August 2019, 16 patients (88% bicuspid aortic valves, no patients with connective tissue disorders) underwent aortic annuloplasty with an external complete Dacron band. Clinical and echocardiographic follow-up was 100% complete. Clinical and echocardiographic follow-up averaged 24.4 ± 9.3 and  15.1 ±  8.3 months, respectively.

Results

Mean cardiopulmonary and cross-clamp times were 105 ± 15 (72-127) and 86 ± 15 (51-113) min, respectively. Early and late mortality was 0%, with no incidents of endocarditis or cerebrovascular events during the follow-up. Two patients were re-operated during the follow-up, one due recurrent aortic regurgitation (12 months after the first operation) yielding a freedom from reoperation due to AR at 1 year and 3 years of 100% ± 0% and 93.3% ± 5.7%, respectively. Based on the latest echocardiogram, five patients had either none or trivial AR, six had mild AR, and three had mild-to-moderate AR.

Conclusions

The early clinical and echocardiographic results after using a complete external Dacron band are promising; however, more data and longer follow-up are needed to determine its role in annular management during aortic valve repair.

SUBMITTER: Kvitting JP 

PROVIDER: S-EPMC8930909 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Early results after aortic annuloplasty with a complete external Dacron band.

Kvitting John-Peder Escobar JP   Beitnes Jan Otto JO   Lundblad Runar R  

General thoracic and cardiovascular surgery 20210920 4


<h4>Objective</h4>This study evaluates the early results of our initial experience with aortic annuloplasty using a complete external Dacron band in the setting of type Ic or type II aortic regurgitation (AR).<h4>Methods</h4>From May 2017 to August 2019, 16 patients (88% bicuspid aortic valves, no patients with connective tissue disorders) underwent aortic annuloplasty with an external complete Dacron band. Clinical and echocardiographic follow-up was 100% complete. Clinical and echocardiographi  ...[more]

Similar Datasets

| S-EPMC10615915 | biostudies-literature
| S-EPMC9367630 | biostudies-literature
| S-EPMC7867416 | biostudies-literature
| S-EPMC5147480 | biostudies-literature
| S-EPMC10193250 | biostudies-literature
| S-EPMC8691863 | biostudies-literature
| S-EPMC10517645 | biostudies-literature
| S-EPMC10405091 | biostudies-literature
| S-EPMC9737026 | biostudies-literature
| S-EPMC10580170 | biostudies-literature