Quantification of Myoinositol in Serum by Electrochemical Detection with an Unmodified Screen-Printed Carbon Electrode.


ABSTRACT: Simple, rapid, and accurate detection of myoinositol (MI) concentration in blood is crucial in diagnosing polycystic ovary syndrome, neurological disorders, and cancer. A novel electrochemical detection (IED) method was established to quantify MI in human serum using a disposable unmodified screen-printed carbon electrode (SPCE) for the first time. MI was detected indirectly by the reaction product of myoinositol dehydrogenase (IDH) and cofactor β-nicotinamide adenine dinucleotide (NAD+). Good linear calibration curves were obtained at the concentration range from 5.0 μM to 500.0 μM (R 2 = 0.9981) with the lower limits of detection (LOD) and quantification (LOQ) of 1.0 μM and 2.5 μM, respectively. Recoveries were calculated at