R-wave amplitude changes with posture and physical activity over time in an insertable cardiac monitor.


ABSTRACT:

Background

Insertable cardiac monitors (ICMs) are accepted tools in cardiac arrhythmia management. Consistent R-wave amplitude (RWA) is essential for optimal detection.

Objectives

Assess RWAs with posture/activities at insertion and at 30 days.

Methods

Participants (n = 90) with Confirm Rx™ ICM had RWAs measured in different postures (supine, right-side [RS], left-side [LS], sitting, and standing) and defined physical activities (including isometric push [IPUSH] and pull) at 2 time points. ICMs were inserted in 45° to sternum and parasternal orientations.

Results

There were significant reductions at insertion with RS, LS, sitting, or standing vs supine (reference position) (all P < .05). At 30 days, significant changes only occurred with LS and sitting (