Combined RT-PCR and Host Response Point-of-Care Testing in Patients Hospitalised with Suspected COVID-19: A Prospective Diagnostic Accuracy Study.


ABSTRACT:

Introduction

RT-PCR has suboptimal sensitivity for the diagnosis of COVID-19. A composite reference standard comprising RT-PCR plus radiological and clinical features has been recommended for diagnostic accuracy studies. The FebriDx finger prick point-of-care test detects an antiviral host response protein (MxA) in 10 min. We evaluated the diagnostic accuracy of FebriDx and RT-PCR compared to a composite reference standard.

Methods

Adults presenting to hospital with suspected COVID-19 were tested by FebriDx and RT-PCR. A composite reference standard was used to classify patients as having COVID-19 based on RT-PCR positivity, or RT-PCR negativity with COVID-19 radiological findings or other clinical criteria. Measures of accuracy were calculated for MxA alone, RT-PCR alone, a