Combined RT-PCR and Host Response Point-of-Care Testing in Patients Hospitalised with Suspected COVID-19: A Prospective Diagnostic Accuracy Study.


ABSTRACT:

Introduction

RT-PCR has suboptimal sensitivity for the diagnosis of COVID-19. A composite reference standard comprising RT-PCR plus radiological and clinical features has been recommended for diagnostic accuracy studies. The FebriDx finger prick point-of-care test detects an antiviral host response protein (MxA) in 10 min. We evaluated the diagnostic accuracy of FebriDx and RT-PCR compared to a composite reference standard.

Methods

Adults presenting to hospital with suspected COVID-19 were tested by FebriDx and RT-PCR. A composite reference standard was used to classify patients as having COVID-19 based on RT-PCR positivity, or RT-PCR negativity with COVID-19 radiological findings or other clinical criteria. Measures of accuracy were calculated for MxA alone, RT-PCR alone, and both combined. This study is registered with the ISRCTN (ISRCTN14966673) and has completed.

Results

A total of 478 patients were tested, with valid results in 475. Of these 475 patients, 222 (46.7%) were classified as having COVID-19; 192 (40.4%) were RT-PCR positive, and 30 (6.3%) were RT-PCR negative and diagnosed on radiological/clinical criteria. Sensitivity of FebriDx MxA vs the composite reference standard was 186/222 (83.8%, 95% CI 78.3-88.4) and was similar to the sensitivity of RT-PCR (192/222 (86.5%, 95% CI 81.3-90.7), (difference of 2.7%, 95% CI - 3.9 to 9.3, p = 0.42). The sensitivity of combined FebriDx and RT-PCR was 208/222 (93.7%) which was superior to both RT-PCR alone (difference of 9.9, 95% CI 4.1-15.9; p = 0.001) and FebriDx MxA alone (difference of 7.2, 95% CI 1.6-12.9; p = 0.011).

Conclusion

Sensitivity of combined FebriDx and RT-PCR testing was superior to each alone for the detection of COVID-19 in hospital and may improve infection control and treatment decisions.

SUBMITTER: Brendish NJ 

PROVIDER: S-EPMC9083481 | biostudies-literature | 2022 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Combined RT-PCR and Host Response Point-of-Care Testing in Patients Hospitalised with Suspected COVID-19: A Prospective Diagnostic Accuracy Study.

Brendish Nathan J NJ   Tanner Alex R AR   Poole Stephen S   Beard Kate R KR   Naidu Vasanth V VV   Mansbridge Christopher T CT   Norton Nicholas J NJ   Wheeler Helen H   Presland Laura L   Clark Tristan W TW  

Infectious diseases and therapy 20220509 3


<h4>Introduction</h4>RT-PCR has suboptimal sensitivity for the diagnosis of COVID-19. A composite reference standard comprising RT-PCR plus radiological and clinical features has been recommended for diagnostic accuracy studies. The FebriDx finger prick point-of-care test detects an antiviral host response protein (MxA) in 10 min. We evaluated the diagnostic accuracy of FebriDx and RT-PCR compared to a composite reference standard.<h4>Methods</h4>Adults presenting to hospital with suspected COVI  ...[more]

Similar Datasets

| S-EPMC7306108 | biostudies-literature
| S-EPMC8263856 | biostudies-literature
| S-EPMC10103696 | biostudies-literature
| S-EPMC7038328 | biostudies-literature
| S-EPMC10562777 | biostudies-literature
| S-EPMC8091859 | biostudies-literature
| S-EPMC9475223 | biostudies-literature
| S-EPMC9081290 | biostudies-literature
2022-03-08 | E-MTAB-11523 | biostudies-arrayexpress
| S-EPMC9222155 | biostudies-literature