ABSTRACT: Background and aims
In late 2019, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) emerged in Wuhan, China. Rapid global spread led to the coronavirus disease 2019 (Covid-19) pandemic. Accurate detection of SARS-CoV-2 has become a vitally important tool in controlling the spread of the virus. Lateral flow devices (LFDs) offer the potential advantage of speed and on-site testing. The sensitivity of these devices compared to reverse transcription-polymerase chain reaction (RT-PCR) has been questioned.Methods
We compared the sensitivity of the Innova LFD, used widely in the United Kingdom, with our rapid RT-PCR method using stored positive samples. Samples with a range of viral loads (original Ct values 18.9-36.5) were tested.Results
The Innova LFD was foun