Investigation of the one-step electrochemical deposition of graphene oxide-doped poly(3,4-ethylenedioxythiophene)-polyphenol oxidase as a dopamine sensor.


ABSTRACT: In this paper, we fabricated poly(3,4-ethylenedioxythiophene) (PEDOT)-graphene oxide-polyphenol oxidase (PEDOT-GO-PPO) as a dopamine sensor. The morphology of PEDOT-GO-PPO was observed using scanning electron microscopy. Cyclic voltammetry was conducted to study the oxidation-reduction characteristics of dopamine. To optimize the pH, potential and limit of detection of dopamine, the amperometric technique was employed. The found limit of detection was 8 × 10-9 M, and the linear range was from 5 × 10-8 to 8.5 × 10-5 M. The Michaelis-Menten constant (K m) was calculated to be 70.34 μM, and the activation energy of the prepared electrode was 32.75 kJ mol-1. The electrode shows no significant change in the interference study. The modified electrode retains up to 80% of its original activity after 2 months. In the future, the biosensor can be used for the quantification of dopamine in human urine samples. The present modified electrode constitutes a tool for the electrochemical analysis of dopamine.

SUBMITTER: Ramu P 

PROVIDER: S-EPMC9125988 | biostudies-literature | 2022 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Investigation of the one-step electrochemical deposition of graphene oxide-doped poly(3,4-ethylenedioxythiophene)-polyphenol oxidase as a dopamine sensor.

Ramu P P   Vimal S P SP   Suresh P P   Sanmugam Anandhavelu A   Saravanakumar U U   Kumar Raju Suresh RS   Almansour Abdulrahman I AI   Arumugam Natarajan N   Vikraman Dhanasekaran D  

RSC advances 20220523 24


In this paper, we fabricated poly(3,4-ethylenedioxythiophene) (PEDOT)-graphene oxide-polyphenol oxidase (PEDOT-GO-PPO) as a dopamine sensor. The morphology of PEDOT-GO-PPO was observed using scanning electron microscopy. Cyclic voltammetry was conducted to study the oxidation-reduction characteristics of dopamine. To optimize the pH, potential and limit of detection of dopamine, the amperometric technique was employed. The found limit of detection was 8 × 10<sup>-9</sup> M, and the linear range  ...[more]

Similar Datasets

| S-EPMC9125232 | biostudies-literature
| S-EPMC9043829 | biostudies-literature
| S-EPMC10188193 | biostudies-literature
| S-EPMC7175412 | biostudies-literature
| S-EPMC9073917 | biostudies-literature
| S-EPMC7658924 | biostudies-literature
| S-EPMC9775058 | biostudies-literature
| S-EPMC9080664 | biostudies-literature
| S-EPMC6780876 | biostudies-literature
| S-EPMC8693727 | biostudies-literature