Impact of acquisition time and misregistration with CT on data-driven gated PET.


ABSTRACT: Objective. Data-driven gating (DDG) can address patient motion issues and enhance PET quantification but suffers from increased image noise from utilization of <100% of PET data. Misregistration between DDG-PET and CT may also occur, altering the potential benefits of gating. Here, the effects of PET acquisition time and CT misregistration were assessed with a combined DDG-PET/DDG-CT technique.Approach. In the primary PET bed with lesions of interest and likely respiratory motion effects, PET acquisition time was extended to 12 min and a low-dose cine CT was acquired to enable DDG-CT. Retrospective reconstructions were created for both non-gated (NG) and DDG-PET using 30 s to 12 min of PET data. Both the standard helical CT and DDG-CT were used for attenuation correction of DDG-PET data. SUVmax, SUVpeak, and CNR were compared for 45 lesions in the liver and lung from 27 cases.Main results. For both NG-PET (p= 0.0041) and DDG-PET (p= 0.0028), only the 30 s acquisition time showed clear SUVmaxbias relative to the 3 min clinical standard. SUVpeakshowed no bias at any change in acquisition time. DDG-PET alone increased SUVmaxby 15 ± 20% (p< 0.0001), then was increased further by an additional 15 ± 29% (p= 0.0007) with DDG-PET/CT. Both 3 min and 6 min DDG-PET had lesion CNR statistically equivalent to 3 min NG-PET, but then increased at 12 min by 28 ± 48% (p= 0.0022). DDG-PET/CT at 6 min had comparable counts to 3 min NG-PET, but significantly increased CNR by 39 ± 46% (p< 0.0001).Significance. 50% counts DDG-PET did not lead to inaccurate or biased SUV-increased SUV resulted from gating. Improved registration from DDG-CT was equally as important as motion correction with DDG-PET for increasing SUV in DDG-PET/CT. Lesion detectability could be significantly improved when DDG-PET used equivalent counts to NG-PET, but only when combined with DDG-CT in DDG-PET/CT.

SUBMITTER: Thomas MA 

PROVIDER: S-EPMC9128538 | biostudies-literature | 2022 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Impact of acquisition time and misregistration with CT on data-driven gated PET.

Thomas M Allan MA   Meier Joseph G JG   Mawlawi Osama R OR   Sun Peng P   Pan Tinsu T  

Physics in medicine and biology 20220408 8


<i>Objective</i>. Data-driven gating (DDG) can address patient motion issues and enhance PET quantification but suffers from increased image noise from utilization of <100% of PET data. Misregistration between DDG-PET and CT may also occur, altering the potential benefits of gating. Here, the effects of PET acquisition time and CT misregistration were assessed with a combined DDG-PET/DDG-CT technique.<i>Approach</i>. In the primary PET bed with lesions of interest and likely respiratory motion e  ...[more]

Similar Datasets

| S-EPMC8403089 | biostudies-literature
| S-EPMC6292629 | biostudies-literature
| S-EPMC6126443 | biostudies-literature
| S-EPMC9399218 | biostudies-literature
| S-EPMC9387080 | biostudies-literature
| S-EPMC8960547 | biostudies-literature
| S-EPMC9529893 | biostudies-literature
| S-EPMC5690042 | biostudies-literature
| S-EPMC7116324 | biostudies-literature
| S-EPMC5734628 | biostudies-literature