Development of Total Lymphoid Irradiation (TLI)-Dedicated Shielding and Image-Guided System and Dose Evaluation Using 3D-Printed Rat Phantom.


ABSTRACT:

Purpose

The purpose of this study is to propose a technique for delivering accurate doses in an image-guided system by developing an experimental setup optimized for total lymphoid irradiation (TLI) in rat lung transplantation.

Materials and methods

In this study, a position-controlled shielding system was developed, and the dose was quantitatively evaluated using a 3D rat phantom and Gafchromic EBT3 film. In addition, we made our own image-guided system that allows the position of the rat and the shielding system to be confirmed during TLI.

Results

As a result of using the position-controlled shielding system, it was found that the doses to the head and lungs were reduced by 93.1 and 87.4%, respectively, of the prescribed doses. In addition, it was shown that the position of the shielding system can be easily confirmed by using the image guidance system.

Conclusion

A shielding apparatus that can control dose delivery according to the size of the rat can optimize the dose for TLI in rat lung transplantation.

SUBMITTER: Choi DH 

PROVIDER: S-EPMC9159376 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Development of Total Lymphoid Irradiation (TLI)-Dedicated Shielding and Image-Guided System and Dose Evaluation Using 3D-Printed Rat Phantom.

Choi Dong Hyeok DH   Ahn So Hyun SH   Park Kwangwoo K   Choi Sang Hyun SH   Kim Jin Sung JS  

Frontiers in veterinary science 20220518


<h4>Purpose</h4>The purpose of this study is to propose a technique for delivering accurate doses in an image-guided system by developing an experimental setup optimized for total lymphoid irradiation (TLI) in rat lung transplantation.<h4>Materials and methods</h4>In this study, a position-controlled shielding system was developed, and the dose was quantitatively evaluated using a 3D rat phantom and Gafchromic EBT3 film. In addition, we made our own image-guided system that allows the position o  ...[more]

Similar Datasets

| S-EPMC9893493 | biostudies-literature
| S-EPMC9079041 | biostudies-literature
| S-EPMC8565655 | biostudies-literature
| S-EPMC11648259 | biostudies-literature
| S-EPMC10659350 | biostudies-literature
| S-EPMC5875554 | biostudies-literature
| S-EPMC6699496 | biostudies-literature
| S-EPMC9327543 | biostudies-literature
| S-EPMC6414136 | biostudies-literature
| S-EPMC6421074 | biostudies-literature