Decomposition of two-component ultrasound pulses in cancellous bone using modified least squares prony method--phantom experiment and simulation.


ABSTRACT: Porous media such as cancellous bone often support the simultaneous propagation of two compressional waves. When small bone samples are interrogated in through-transmission with broadband sources, these two waves often overlap in time. The modified least-squares Prony's (MLSP) method was tested for decomposing a 500 kHz-center-frequency signal containing two overlapping components: one passing through a polycarbonate plate (to produce the "fast" wave) and another passing through a cancellous-bone-mimicking phantom (to produce the "slow" wave). The MLSP method yielded estimates of attenuation slopes accurate to within 7% (polycarbonate plate) and 2% (cancellous bone phantom). The MLSP method yielded estimates of phase velocities accurate to within 1.5% (both media). The MLSP method was also