Managing argon interference during measurements of 18O/16O ratios in O2 by continuous-flow isotope ratio mass spectrometry.


ABSTRACT: Monitoring changes in stable oxygen isotope ratios in molecular oxygen allows for studying many fundamental processes in bio(geo)chemistry and environmental sciences. While the measurement of [Formula: see text]O/[Formula: see text]O ratios of [Formula: see text] in gaseous samples can be carried out conveniently and from extracting moderately small aqueous samples for analyses by continuous-flow isotope ratio mass spectrometry (CF-IRMS), oxygen isotope signatures, [Formula: see text]O, could be overestimated by more than 6[Formula: see text] because of interferences from argon in air. Here, we systematically evaluated the extent of such Ar interferences on [Formula: see text]O/[Formula: see text]O ratios of [Formula: see text] for measurements by gas chromatography/IRMS and GasBench/IRMS and propose simple instrumental modifications for improved Ar and [Formula: see text] separation as well as post-measurement correction procedures for obtaining accurate [Formula: see text]O. We subsequently evaluated the consequences of Ar interferences for the quantification of O isotope fractionation in terms of isotope enrichment factors, [Formula: see text], and [Formula: see text]O kinetic isotope effects ([Formula: see text]O KIEs) in samples where [Formula: see text] is consumed and Ar:[Formula: see text] ratios increase steadily and substantially over the course of a reaction. We show that the extent of O isotope fractionation is overestimated only slightly and that this effect is typically smaller than uncertainties originating from the precision of [Formula: see text]O measurements and experimental variability. Ar interferences can become more relevant and bias [Formula: see text] values by more than [Formula: see text] in aqueous samples where fractional [Formula: see text] conversion exceeds 90%. Practically, however, such samples would typically contain less than 25 [Formula: see text]M of [Formula: see text] at ambient temperature, an amount that is close to the method detection limit of [Formula: see text]O/[Formula: see text]O ratio measurement by CF-IRMS.

SUBMITTER: Bopp CE 

PROVIDER: S-EPMC9314310 | biostudies-literature | 2022 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Managing argon interference during measurements of <sup>18</sup>O/<sup>16</sup>O ratios in O<sub>2</sub> by continuous-flow isotope ratio mass spectrometry.

Bopp Charlotte E CE   Bolotin Jakov J   Pati Sarah G SG   Hofstetter Thomas B TB  

Analytical and bioanalytical chemistry 20220716 20


Monitoring changes in stable oxygen isotope ratios in molecular oxygen allows for studying many fundamental processes in bio(geo)chemistry and environmental sciences. While the measurement of [Formula: see text]O/[Formula: see text]O ratios of [Formula: see text] in gaseous samples can be carried out conveniently and from extracting moderately small aqueous samples for analyses by continuous-flow isotope ratio mass spectrometry (CF-IRMS), oxygen isotope signatures, [Formula: see text]O, could be  ...[more]

Similar Datasets

| S-EPMC3479679 | biostudies-literature
| S-EPMC6082370 | biostudies-literature
| S-EPMC9795564 | biostudies-literature
| S-EPMC10844122 | biostudies-literature
| S-EPMC5881069 | biostudies-literature
| S-EPMC8450417 | biostudies-literature
| S-EPMC6590658 | biostudies-literature
| S-EPMC10638474 | biostudies-literature
| S-EPMC6246559 | biostudies-literature
| S-EPMC5073312 | biostudies-literature