Ontology highlight
ABSTRACT: Objectives
We aimed to develop simple and rapid HPLC methods for determination of amoxicillin and clindamycin in human plasma.Methods
Plasma samples were pretreated by direct deproteinization with acetonitrile and the analytical separation took place on a reverse phase Poroshell 120 EC-C18 column (2.7 μm, 2.1 × 100 mm) with a gradient of acetonitrile. UV detection at 229 nm for amoxicillin and 204 nm for clindamycin was used for determination of the antibiotics in plasma.Results
The calibration curves were linear over the concentration ranges of 1-100 mg/L for amoxicillin and 1-15 mg/L for clindamycin with a correlation coefficient of ≥0.98. Intra-assay precisions were all ≤15% and the accuracies were within ±15%. The limit of quantification (LOQ) was found to be 0.5 mg/L for amoxicillin and 1 mg/L for clindamycin with inter-assay imprecision coefficient of variances (CVs) of 18.7% and 15.6%, respectively. The present HPLC methods were successfully applied on spike-in samples and on plasma samples collected 4-6 and 3.5-5.5 h after oral antibiotic administration of 500 mg of amoxicillin and 600 mg of clindamycin, respectively.Conclusions
We have developed HPLC methods with UV detection for quantification of amoxicillin and clindamycin in human plasma. The methods are fast, simple and suitable for use in routine settings and clinical studies.
SUBMITTER: Greibe E
PROVIDER: S-EPMC9410666 | biostudies-literature | 2022 Aug
REPOSITORIES: biostudies-literature
Greibe Eva E Moser Claus Ernst CE Bruun Niels Eske NE Hoffmann-Lücke Elke E
The Journal of antimicrobial chemotherapy 20220801 9
<h4>Objectives</h4>We aimed to develop simple and rapid HPLC methods for determination of amoxicillin and clindamycin in human plasma.<h4>Methods</h4>Plasma samples were pretreated by direct deproteinization with acetonitrile and the analytical separation took place on a reverse phase Poroshell 120 EC-C18 column (2.7 μm, 2.1 × 100 mm) with a gradient of acetonitrile. UV detection at 229 nm for amoxicillin and 204 nm for clindamycin was used for determination of the antibiotics in plasma.<h4>Resu ...[more]