Rapid Adaptation of Established High-Throughput Molecular Testing Infrastructure for Monkeypox Virus Detection.


ABSTRACT: Beginning in May 2022, a rising number of monkeypox cases were reported in non-monkeypox-endemic countries in the Northern Hemisphere. We adapted 2 published quantitative PCRs for use as a dual-target monkeypox virus test on widely used automated high-throughput PCR systems. We determined analytic performance by serial dilutions of monkeypox virus reference material, which we quantified by digital PCR. We found the lower limit of detection for the combined assays was 4.795 (95% CI 3.6-8.6) copies/mL. We compared clinical performance against a commercial manual orthopoxvirus research use only PCR kit by using clinical remnant swab samples. Our assay showed 100% positive (n = 11) and 100% negative (n = 56) agreement. Timely and scalable PCR tests are crucial for limiting further spread of monkeypox. The assay we provide streamlines high-throughput molecular testing for monkeypox virus on existing broadly established platforms used for SARS-CoV-2 diagnostic testing.

SUBMITTER: Norz D 

PROVIDER: S-EPMC9423910 | biostudies-literature | 2022 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Rapid Adaptation of Established High-Throughput Molecular Testing Infrastructure for Monkeypox Virus Detection.

Nörz Dominik D   Tang Hui Ting HT   Emmerich Petra P   Giersch Katja K   Fischer Nicole N   Schmiedel Stephan S   Addo Marylyn M MM   Aepfelbacher Martin M   Pfefferle Susanne S   Lütgehetmann Marc M  

Emerging infectious diseases 20220729 9


Beginning in May 2022, a rising number of monkeypox cases were reported in non-monkeypox-endemic countries in the Northern Hemisphere. We adapted 2 published quantitative PCRs for use as a dual-target monkeypox virus test on widely used automated high-throughput PCR systems. We determined analytic performance by serial dilutions of monkeypox virus reference material, which we quantified by digital PCR. We found the lower limit of detection for the combined assays was 4.795 (95% CI 3.6-8.6) copie  ...[more]

Similar Datasets

| S-EPMC9573898 | biostudies-literature
| S-EPMC9629024 | biostudies-literature
| S-EPMC10335395 | biostudies-literature
| S-EPMC9792191 | biostudies-literature
| S-EPMC11496496 | biostudies-literature
| S-EPMC9603857 | biostudies-literature
| S-EPMC5146750 | biostudies-literature
| S-EPMC10080115 | biostudies-literature