ABSTRACT: Purpose
To evaluate the safety and effectiveness of the CT LUCIA 611P intraocular lens (IOL) in patients with cataracts.Setting
23 surgeons at 15 different clinical sites.Design
Prospective single-arm clinical trial.Methods
The study was conducted under an Investigational Device Exemption for premarket approval of a new hydrophobic acrylic IOL in the United States. Patients were followed for 12 months, and the main measured variables included uncorrected (UDVA) and corrected (CDVA) distance visual acuities, manifest refraction, and adverse events.Results
In total, 339 eyes of 339 patients were implanted with the study device, of which 310 (91.4%) reached the 12-month visit. The percentage of eyes within ±0.50 diopter (D) and ±1.00 D of emmetropia was 85.8% (266/310) and 96.8% (300/310), respectively. Manifest refraction spherical equivalent (MRSE) remained stable over the first year with the mean 12-month MRSE of -0.03 ± 0.45 D. The mean 12-month UDVA and CDVA were 0.09 ± 0.15 (≈20/25) and -0.02 ± 0.09 (≈20/19) logMAR, respectively. Of all patients, 99.4% (308/310) achieved postoperative CDVA ≥20/40. The incidence of Nd:YAG capsulotomy within the first year was 3.5% (11/310). Only 2 eyes had IOL tilt present at the 12-month postoperative visit with no associated visual symptoms. There were 2 cases of IOL decentration; one required removal of the IOL, whereas the other had no visual side effects related to decentration. There were no findings of glistening at any visit.Conclusions
The CT LUCIA 611P IOL demonstrated excellent safety, efficacy, and stability of refractive outcomes. No significant issues related to the biocompatibility of the IOL material were observed.
SUBMITTER: Schallhorn SC
PROVIDER: S-EPMC9514744 | biostudies-literature | 2022 Oct
REPOSITORIES: biostudies-literature
Schallhorn Steven C SC Bonilla Michael M Pantanelli Seth M SM
Journal of cataract and refractive surgery 20220325 10
<h4>Purpose</h4>To evaluate the safety and effectiveness of the CT LUCIA 611P intraocular lens (IOL) in patients with cataracts.<h4>Setting</h4>23 surgeons at 15 different clinical sites.<h4>Design</h4>Prospective single-arm clinical trial.<h4>Methods</h4>The study was conducted under an Investigational Device Exemption for premarket approval of a new hydrophobic acrylic IOL in the United States. Patients were followed for 12 months, and the main measured variables included uncorrected (UDVA) an ...[more]