Systolic Arterial Pressure Control Using an Automated Closed-Loop System for Vasopressor Infusion during Intermediate-to-High-Risk Surgery: A Feasibility Study.


ABSTRACT: Introduction: Vasopressor infusions are essential in treating and preventing intraoperative hypotension. Closed-loop vasopressor therapy outperforms clinicians when the target is set at a mean arterial pressure (MAP) baseline, but little is known on the performance metrics of closed-loop vasopressor infusions when systolic arterial pressure (SAP) is the controlled variable. Methods: Patients undergoing intermediate- to high-risk abdominal surgery were included in this prospective cohort feasibility study. All patients received norepinephrine infusion through a computer controlled closed-loop system that targeted SAP at 130 mmHg. The primary objective was to determine the percent of case time in hypotension or under target defined as SAP below 10% of the target (SAP < 117 mmHg). Secondary o