Integral assessment of gas exchange during veno-arterial ECMO: accuracy and precision of a modified Fick principle in a porcine model.
Ontology highlight
ABSTRACT: Assessment of native cardiac output during extracorporeal circulation is challenging. We assessed a modified Fick principle under conditions such as dead space and shunt in 13 anesthetized swine undergoing centrally cannulated veno-arterial extracorporeal membrane oxygenation (V-A ECMO, 308 measurement periods) therapy. We assumed that the ratio of carbon dioxide elimination (V̇co2) or oxygen uptake (V̇o2) between the membrane and native lung corresponds to the ratio of respective blood flows. Unequal ventilation/perfusion (V̇/Q̇) ratios were corrected towards unity. Pulmonary blood flow was calculated and compared to an ultrasonic flow probe on the pulmonary artery with a bias of 99 mL/min (limits of agreement -542 to 741 mL/min) with blood content V̇o2 an
SUBMITTER: Berger DC
PROVIDER: S-EPMC9870575 | biostudies-literature | 2023 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA