Safe and effective delivery of high-power, short-duration radiofrequency ablation lesions with a flexible-tip ablation catheter.


ABSTRACT:

Background

High-power, short-duration (HPSD) radiofrequency ablation (RFA) may reduce ablation time. Concerns that catheter-mounted thermocouples (TCs) can underestimate tissue temperature, resulting in elevated risk of steam pop formation, potentially limit widespread adoption of HPSD ablation.

Objective

The purpose of this study was to compare the safety and efficacy of HPSD and low-power, long-duration (LPLD) RFA in the context of pulmonary vein isolation (PVI).

Methods

An open-irrigated ablation catheter with a contact force sensor and a flexible-tip electrode containing a TC at its distal end (TactiFlexTM Ablation Catheter, Sensor EnabledTM, Abbott) was used to isolate the left pulmonary veins (PVs) in 12 canines with HPSD RFA (50 W for 10 s

SUBMITTER: Ptaszek LM 

PROVIDER: S-EPMC9877396 | biostudies-literature | 2023 Jan