"Real-world" analysis of battery longevity of implantable cardioverter-defibrillators: an in-depth analysis of a prospective defibrillator database.


ABSTRACT:

Background

There is a lack of evidence regarding contemporary implantable cardioverter-defibrillator (ICD) battery longevity. Our aim was to assess battery longevity in ICDs in a real-world setting.

Methods

Retrospective cross-sectional single center study of a prospectively collected database of consecutive patients who underwent ICD implantation from January 2010 to December 2015. Clinical data and battery longevity of all manufacturers were collected.

Results

A total of 351 patients (84.6% males, mean age of 61 ± 12 years) were included in the study (292 VVI; 6 VDD; 53 DDD). All manufacturers (Abbott, Biotronik, Boston, Medtronic and Microport) were equally represented in the study (p = 0.110). Median battery longevity was 10.8 years (11 years for VVI and 8.5 for D