DeepSMCP - Deep-learning powered denoising of Monte Carlo dose distributions within the Swiss Monte Carlo Plan.


ABSTRACT: This work demonstrated the development of a fast, deep-learning framework (DeepSMCP) to mitigate noise in Monte Carlo dose distributions (MC-DDs) of photon treatment plans with high statistical uncertainty (SU) and its integration into the Swiss Monte Carlo Plan (SMCP). To this end, a two-channel input (MC-DD and computed tomography (CT) scan) 3D U-net was trained, validated and tested (80%/10%/10%) on high/low-SU MC-DD-pairs of 106 clinically-motivated VMAT arcs for 29 available CTs, augmented to 3074 pairs. The model was integrated into SMCP to enable a "one-click" workflow of calculating and denoising MC-DDs of high SU to obtain MC-DDs of low SU. The model accuracy was evaluated on the test set using Gamma passing rate (2% global, 2 mm, 10% threshold) comparing denoised and low-SU MC-DD. Calculation time for the whole workflow was recorded. Denoised MC-DDs match low-SU MC-DDs with average (standard deviation) Gamma passing rate of 82.9% (4.7%). Additional application of DeepSMCP to 12 unseen clinically-motivated cases of different treatment sites, including treatment sites not present during training, resulted in an average Gamma passing rate of 91.0%. Denoised DDs were obtained on average in 35.1 s, a 340-fold efficiency gain compared to low-SU MC-DD calculation. DeepSMCP presented a first seamlessly integrated promising denoising framework for MC-DDs.

SUBMITTER: Loebner HA 

PROVIDER: S-EPMC12901522 | biostudies-literature | 2026 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

DeepSMCP - Deep-learning powered denoising of Monte Carlo dose distributions within the Swiss Monte Carlo Plan.

Loebner Hannes A HA   Joost Raphael R   Bertholet Jenny J   Mougiakakou Stavroula S   Fix Michael K MK   Manser Peter P  

Zeitschrift fur medizinische Physik 20250317 1


This work demonstrated the development of a fast, deep-learning framework (DeepSMCP) to mitigate noise in Monte Carlo dose distributions (MC-DDs) of photon treatment plans with high statistical uncertainty (SU) and its integration into the Swiss Monte Carlo Plan (SMCP). To this end, a two-channel input (MC-DD and computed tomography (CT) scan) 3D U-net was trained, validated and tested (80%/10%/10%) on high/low-SU MC-DD-pairs of 106 clinically-motivated VMAT arcs for 29 available CTs, augmented  ...[more]

Similar Datasets

| S-EPMC9127420 | biostudies-literature
| S-EPMC5875513 | biostudies-literature
| S-EPMC11929889 | biostudies-literature
| S-EPMC11524488 | biostudies-literature
| S-EPMC9845370 | biostudies-literature
| S-EPMC5722431 | biostudies-literature
| S-EPMC8833269 | biostudies-literature
| S-EPMC9674698 | biostudies-literature
| S-EPMC5718669 | biostudies-literature
| S-EPMC8134502 | biostudies-literature