Closed-Loop Control Performance of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System in a Feasibility Study.


ABSTRACT:

Background

This feasibility study investigated the insulin-delivery characteristics of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System-an automated insulin delivery device-in participants with type 1 diabetes.

Methods

Thirteen adults with type 1 diabetes were enrolled in this nonrandomized, uncontrolled, clinical-research-center-based feasibility study. The HHM System comprised a continuous subcutaneous insulin infusion pump, a continuous glucose monitor (CGM), and a model predictive control algorithm with a safety module, run on a laptop platform. Closed-loop control lasted approximately 20 hours, including an overnight period and two meals.

Results

When attempting to minimize glucose excursions outside of a prespecified target zone, the predictive HHM System d