ABSTRACT: Background
This feasibility study investigated the insulin-delivery characteristics of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System-an automated insulin delivery device-in participants with type 1 diabetes.Methods
Thirteen adults with type 1 diabetes were enrolled in this nonrandomized, uncontrolled, clinical-research-center-based feasibility study. The HHM System comprised a continuous subcutaneous insulin infusion pump, a continuous glucose monitor (CGM), and a model predictive control algorithm with a safety module, run on a laptop platform. Closed-loop control lasted approximately 20 hours, including an overnight period and two meals.Results
When attempting to minimize glucose excursions outside of a prespecified target zone, the predictive HHM System decreased insulin infusion rates below the participants' preset basal rates in advance of below-zone excursions (CGM < 90 mg/dl), and delivered 80.4% less insulin than basal during those excursions. Similarly, the HHM System increased infusion rates above basal during above-zone excursions (CGM > 140 mg/dl), delivering 39.9% more insulin than basal during those excursions. Based on YSI, participants spent a mean ± standard deviation (SD) of 0.2 ± 0.5% of the closed-loop control time at glucose levels < 70 mg/dl, including 0.3 ± 0.9% for the overnight period. The mean ± SD glucose based on YSI for all participants was 164.5 ± 23.5 mg/dl. There were nine instances of algorithm-recommended supplemental carbohydrate administrations, and there was no severe hypoglycemia or diabetic ketoacidosis.Conclusions
Results of this study indicate that the current HHM System is a feasible foundation for development of a closed-loop insulin delivery device.
SUBMITTER: Finan DA
PROVIDER: S-EPMC4454115 | biostudies-literature | 2014 Jan
REPOSITORIES: biostudies-literature
Finan Daniel A DA McCann Thomas W TW Mackowiak Linda L Dassau Eyal E Patek Stephen D SD Kovatchev Boris P BP Doyle Francis J FJ Zisser Howard H Anhalt Henry H Venugopalan Ramakrishna R
Journal of diabetes science and technology 20140101 1
<h4>Background</h4>This feasibility study investigated the insulin-delivery characteristics of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System-an automated insulin delivery device-in participants with type 1 diabetes.<h4>Methods</h4>Thirteen adults with type 1 diabetes were enrolled in this nonrandomized, uncontrolled, clinical-research-center-based feasibility study. The HHM System comprised a continuous subcutaneous insulin infusion pump, a continuous glucose monitor (CGM), and a model p ...[more]