B0 and B1 inhomogeneities in the liver at 1.5 T and 3.0 T.


ABSTRACT:

Purpose

The purpose of this work is to characterize the magnitude and variability of B0 and B1 inhomogeneities in the liver in large cohorts of patients at both 1.5 T and 3.0 T.

Methods

Volumetric B0 and B1 maps were acquired over the liver of patients presenting for routine abdominal MRI. Regions of interest were drawn in the nine Couinaud segments of the liver, and the average value was recorded. Magnitude and variation of measured averages in each segment were reported across all patients.

Results

A total of 316 B0 maps and 314 B1 maps, acquired at 1.5 T and 3.0 T on a variety of GE Healthcare MRI systems in 630 unique exams, were identified, analyzed, and, in the interest of reproducible research, de-identified and made public. Measured B0 inhomogeneities ranged (5th-95th percentiles) from -31.7 Hz to 164.0 Hz for 3.0 T (-14.5 Hz to 81.3 Hz at 1.5 T), while measured B1 inhomogeneities (ratio of actual over prescribed flip angle) ranged from 0.59 to 1.13 for 3.0 T (0.83 to 1.11 at 1.5 T).

Conclusion

This study provides robust characterization of B0 and B1 inhomogeneities in the liver to guide the development of imaging applications and protocols. Field strength, bore diameter, and sex were determined to be statistically significant effects for both B0 and B1 uniformity. Typical clinical liver imaging at 3.0 T should expect B0 inhomogeneities ranging from approximately -100 Hz to 250 Hz (-50 Hz to 150 Hz at 1.5 T) and B1 inhomogeneities ranging from approximately 0.4 to 1.3 (0.7 to 1.2 at 1.5 T).

SUBMITTER: Roberts NT 

PROVIDER: S-EPMC7869971 | biostudies-literature | 2021 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

B<sub>0</sub> and B<sub>1</sub> inhomogeneities in the liver at 1.5 T and 3.0 T.

Roberts Nathan T NT   Hinshaw Louis A LA   Colgan Timothy J TJ   Ii Takanori T   Hernando Diego D   Reeder Scott B SB  

Magnetic resonance in medicine 20201026 4


<h4>Purpose</h4>The purpose of this work is to characterize the magnitude and variability of B<sub>0</sub> and B<sub>1</sub> inhomogeneities in the liver in large cohorts of patients at both 1.5 T and 3.0 T.<h4>Methods</h4>Volumetric B<sub>0</sub> and B<sub>1</sub> maps were acquired over the liver of patients presenting for routine abdominal MRI. Regions of interest were drawn in the nine Couinaud segments of the liver, and the average value was recorded. Magnitude and variation of measured ave  ...[more]

Similar Datasets

| S-EPMC9885339 | biostudies-literature
| S-EPMC9311141 | biostudies-literature
| S-EPMC11502507 | biostudies-literature
| S-EPMC9892291 | biostudies-literature
| S-EPMC7683379 | biostudies-literature
| S-EPMC10972246 | biostudies-literature
| S-EPMC9875455 | biostudies-literature
| S-EPMC10619883 | biostudies-literature
| S-EPMC2704948 | biostudies-other
| S-EPMC8624877 | biostudies-literature