Assessment of TMT Labeling Efficiency in Large-Scale Quantitative Proteomics: The Critical Effect of Sample pH.


ABSTRACT: Isobaric labeling via tandem mass tag (TMT) reagents enables sample multiplexing prior to LC-MS/MS, facilitating high-throughput large-scale quantitative proteomics. Consistent and efficient labeling reactions are essential to achieve robust quantification; therefore, embedded in our clinical proteomic protocol is a quality control (QC) sample that contains a small aliquot from each sample within a TMT set, referred to as "Mixing QC." This Mixing QC enables the detection of TMT labeling issues by LC-MS/MS before combining the full samples to allow for salvaging of poor TMT labeling reactions. While TMT labeling is a valuable tool, factors leading to poor reactions are not fully studied. We observed that relabeling does not necessarily rescue TMT reactions and that peptide samples sometimes

SUBMITTER: Hutchinson-Bunch C 

PROVIDER: S-EPMC8154127 | biostudies-literature | 2021 May