Oxygen-enhanced functional lung imaging using a contemporary 0.55 T MRI system.


ABSTRACT: The purpose of this study was to evaluate oxygen-enhanced pulmonary imaging at 0.55 T with 3D stack-of-spirals ultrashort-TE (UTE) acquisition. Oxygen-enhanced pulmonary MRI offers the measurement of regional lung ventilation and perfusion using inhaled oxygen as a contrast agent. Low-field MRI systems equipped with contemporary hardware can provide high-quality structural lung imaging by virtue of the prolonged T2 *. Fortuitously, the T1 relaxivity of oxygen increases at lower field strengths, which is expected to improve the sensitivity of oxygen-enhanced lung MRI. We implemented a breath-held T1 -weighted 3D stack-of-spirals UTE acquisition with a 7 ms spiral-out readout. Measurement repeatability was assessed using five repetitions of oxygen-enhanced lung imaging in healthy volunteers (n = 7). The signal intensity at both normoxia and hyperoxia was strongly dependent on lung tissue density modulated by breath-hold volume during the five repetitions. A voxel-wise correction for lung tissue density improved the repeatability of percent signal enhancement maps (coefficient of variation = 34 ± 16%). Percent signal enhancement maps were compared in 15 healthy volunteers and 10 patients with lymphangioleiomyomatosis (LAM), a rare cystic disease known to reduce pulmonary function. We measured a mean percent signal enhancement of 9.0 ± 3.5% at 0.55 T in healthy volunteers, and reduced signal enhancement in patients with LAM (5.4 ± 4.8%, p = 0.02). The heterogeneity, estimated by the percent of lung volume exhibiting low enhancement, was significantly increased in patients with LAM compared with healthy volunteers (11.1 ± 6.0% versus 30.5 ± 13.1%, p = 0.01), illustrating the capability to measure regional functional deficits.

SUBMITTER: Bhattacharya I 

PROVIDER: S-EPMC8377594 | biostudies-literature | 2021 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications


The purpose of this study was to evaluate oxygen-enhanced pulmonary imaging at 0.55 T with 3D stack-of-spirals ultrashort-T<sub>E</sub> (UTE) acquisition. Oxygen-enhanced pulmonary MRI offers the measurement of regional lung ventilation and perfusion using inhaled oxygen as a contrast agent. Low-field MRI systems equipped with contemporary hardware can provide high-quality structural lung imaging by virtue of the prolonged T<sub>2</sub> *. Fortuitously, the T<sub>1</sub> relaxivity of oxygen inc  ...[more]

Similar Datasets

| S-EPMC10947575 | biostudies-literature
| S-EPMC8259660 | biostudies-literature
| S-EPMC9018883 | biostudies-literature
| S-EPMC11421172 | biostudies-literature
| S-EPMC10952250 | biostudies-literature
| S-EPMC6312433 | biostudies-literature
| S-EPMC8749356 | biostudies-literature
| S-EPMC6295484 | biostudies-literature
| S-EPMC7856701 | biostudies-literature
| S-EPMC5899670 | biostudies-literature