Technical note: Partitioning of gated single photon emission computed tomography raw data for protocols optimization.


ABSTRACT:

Purpose

Methodologies for optimization of SPECT image acquisition can be challenging due to imaging throughput, physiological bias, and patient comfort constraints. We evaluated a vendor-independent method for simulating lower count image acquisitions.

Methods

We developed an algorithm that recombines the ECG-gated raw data into reduced counting acquisitions. We then tested the algorithm to simulate reduction of counting statistics from phantom SPECT image acquisition, which was synchronized with an ECG simulator. The datasets were reconstructed with a resolution recovery algorithm and the summed stress score (SSS) was assessed by three readers (two experts and one automatic).

Results

The algorithm generated varying counting levels, simulating multiple examinations at the same time. The error between the expected and the simulated countings ranged from approximately 5% to 10% for the ungated simulations and 0% for the gated simulations.

Conclusions

The vendor-independent algorithm successfully generated lower counting statistics datasets from single-gated SPECT raw data. This method can be readily implemented for optimal SPECT research aiming to lower the injected activity and/ or to shorten the acquisition time.

SUBMITTER: Queiroz CC 

PROVIDER: S-EPMC8906212 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Technical note: Partitioning of gated single photon emission computed tomography raw data for protocols optimization.

Queiroz Cleiton Cavalcante CC   Machado Marcos Antonio Dorea MAD   Ximenes Antonio Augusto Brito AAB   Pino Andre Gustavo Silva AGS   Netto Eduardo Martins EM  

Journal of applied clinical medical physics 20211217 3


<h4>Purpose</h4>Methodologies for optimization of SPECT image acquisition can be challenging due to imaging throughput, physiological bias, and patient comfort constraints. We evaluated a vendor-independent method for simulating lower count image acquisitions.<h4>Methods</h4>We developed an algorithm that recombines the ECG-gated raw data into reduced counting acquisitions. We then tested the algorithm to simulate reduction of counting statistics from phantom SPECT image acquisition, which was s  ...[more]

Similar Datasets

| S-EPMC6916323 | biostudies-literature
| S-EPMC5721092 | biostudies-literature
| S-EPMC3358562 | biostudies-literature
| S-EPMC5578892 | biostudies-literature
| S-EPMC6283982 | biostudies-literature
| S-EPMC6558785 | biostudies-literature
| S-EPMC10958524 | biostudies-literature
| S-EPMC7475100 | biostudies-literature
| S-EPMC5473361 | biostudies-literature
| S-EPMC5058868 | biostudies-other