ABSTRACT: Purpose
To present a new complex-valued B1+ mapping method for electrical properties tomography using Carr-Purcell spin echoes.Methods
A Carr-Purcell (CP) echo train generates pronounced flip-angle dependent oscillations that can be used to estimate the magnitude of B1+ . To this end, a dictionary is used that takes into account the slice profile as well as T2 relaxation along the echo train. For validation, the retrieved B1+ map is compared with the actual flip angle imaging (AFI) method in a phantom (79 ε0 , 0.34 S/m). Moreover, the phase of the first echo reflects the transceive phase. Overall, the CP echo train yields an estimate of the complex-valued B1+ , allowing electrical properties tomography with both permittivity and conductivity. The presented method is evaluated in phantom scans as well as for in vivo brain at 3 T.Results
In the phantom, the obtained magnitude B1+ maps retrieved from the CP echo train and the AFI method show excellent agreement, and both the reconstructed estimated permittivity (79 ± 3) ε0 and conductivity (0.35 ± 0.04) S/m values are in accordance with expectations. In the brain, the obtained electrical properties are also close to expectations. In addition to the retrieved complex B1+ information, the decay of the CP echo trains also yields an estimate for T2 .Conclusion
The CP sequence can be used to simultaneously provide both B1+ magnitude and phase estimations, and therefore allows for full reconstruction of the electrical properties.
SUBMITTER: Iyyakkunnel S
PROVIDER: S-EPMC9298742 | biostudies-literature | 2022 Mar
REPOSITORIES: biostudies-literature
Iyyakkunnel Santhosh S Weigel Matthias M Ganter Carl C Bieri Oliver O
Magnetic resonance in medicine 20211109 3
<h4>Purpose</h4>To present a new complex-valued B<sub>1</sub><sup>+</sup> mapping method for electrical properties tomography using Carr-Purcell spin echoes.<h4>Methods</h4>A Carr-Purcell (CP) echo train generates pronounced flip-angle dependent oscillations that can be used to estimate the magnitude of B<sub>1</sub><sup>+</sup> . To this end, a dictionary is used that takes into account the slice profile as well as T<sub>2</sub> relaxation along the echo train. For validation, the retrieved B<s ...[more]