Repeatability, reproducibility, and accuracy of quantitative mri of the breast in the community radiology setting.
Ontology highlight
ABSTRACT: Quantitative diffusion-weighted MRI (DW-MRI) and dynamic contrast-enhanced MRI (DCE-MRI) have the potential to impact patient care by providing noninvasive biological information in breast cancer. To quantify the repeatability, reproducibility, and accuracy of apparent diffusion coefficient (ADC) and T1 -mapping of the breast in community radiology practices. Prospective. Ice-water DW-MRI and T1 gel phantoms were used to assess accuracy. Normal subjects (n = 3) and phantoms across three sites (one academic, two community) were used to assess reproducibility. Test-retest analysis at one site in normal subjects (n = 12) was used to assess repeatability. 3T Siemens Skyra MRI quantitative DW-MRI and T1 -mapping. Quantitative DW-MRI and T1 -mapping parametric maps of phantoms and fibroglandular and adipose tissue of the breast. Average values of breast tissue were quantified and Bland-Altman analysis was performed to assess the repeatability of the MRI techniques, while the Friedman test assessed reproducibility. ADC measurements were reproducible across sites, with an average difference of 1.6% in an ice-water phantom and 7.0% in breast fibroglandular tissue. T1 measurements in gel phantoms had an average difference of 2.8% across three sites, whereas breast fibroglandular and adipose tissue had 8.4% and 7.5% average differences, respectively. In the repeatability study, we found no bias between first and second scanning sessions (P = 0.1). The difference between repeated measurements was independent of the mean for each MRI metric (P = 0.156, P = 0.862, P = 0.197 for ADC, T1 of fibroglandular tissue, and T1 of adipose tissue, respectively). Community radiology practices can perform repeatable, reproducible, and accurate quantitative T1 -mapping and DW-MRI. This has the potential to dramatically expand the number of sites that can participate in multisite clinical trials and increase clinical translation of quantitative MRI techniques for cancer response assessment. 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2018.
SUBMITTER: Sorace AG
PROVIDER: S-EPMC6151298 | biostudies-literature | 2018 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA