Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system for 18F-FDG.


ABSTRACT:

Aim

Recently, total-body PET/CT systems with an extended axial field-of-view (aFOV) became commercially available which allow acquiring physiologic information of multiple organs simultaneously. However, the nominal aFOV may clinically not be used effectively due to the inherently reduced sensitivity at the distal ends of the aFOV. The aim of this study was to assess the extent of the useful aFOV of the Biograph Vision Quadra PET/CT system.

Methods

A NEMA image quality (IQ) phantom mimicking a standard [18F]FDG examination was used. Image contrast and noise were assessed across the 106 cm aFOV of the Biograph Vision Quadra PET/CT system (Siemens Healthineers). Phantom acquisitions were performed at different axial positions. PET data were rebinned to simulate different acquisition times for a standard injected activity and reconstructed using different filter settings to evaluate the noise and images along the axial direction.

Results

Image noise and contrast were stable within the central 80 cm of the aFOV. Outside this central area, image contrast variability as well as image noise increased. This degradation of IQ was in particular evident for short acquisition times of less than 30 s. At 10 min acquisition time and in the absence of post-reconstruction filtering, the useful aFOV was 100 cm. For a 2 min acquisition time, a useful aFOV with image noise below 15% was only achievable using Gaussian filtering with axial extents of between 83 and 103 cm when going from 2 to 6 mm full-width-half-maximum, respectively.

Conclusion

Image noise increases substantially towards the ends of the aFOV. However, good IQ in compliance with generally accepted benchmarks is achievable for an aFOV of > 90 cm. When accepting higher image noise or using dedicated protocol settings such as stronger filtering a useful aFOV of around 1 m can be achieved for a 2 min acquisition time.

SUBMITTER: Rausch I 

PROVIDER: S-EPMC9747988 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system for <sup>18</sup>F-FDG.

Rausch Ivo I   Mannheim Julia G JG   Kupferschläger Jürgen J   la Fougère Christian C   Schmidt Fabian P FP  

EJNMMI physics 20221214 1


<h4>Aim</h4>Recently, total-body PET/CT systems with an extended axial field-of-view (aFOV) became commercially available which allow acquiring physiologic information of multiple organs simultaneously. However, the nominal aFOV may clinically not be used effectively due to the inherently reduced sensitivity at the distal ends of the aFOV. The aim of this study was to assess the extent of the useful aFOV of the Biograph Vision Quadra PET/CT system.<h4>Methods</h4>A NEMA image quality (IQ) phanto  ...[more]

Similar Datasets

| S-EPMC10224893 | biostudies-literature
| S-EPMC9618000 | biostudies-literature
| S-EPMC7807699 | biostudies-literature
| S-EPMC9931831 | biostudies-literature
| S-EPMC9606094 | biostudies-literature
| S-EPMC10606613 | biostudies-literature
| S-EPMC10611821 | biostudies-literature
| S-EPMC9445107 | biostudies-literature
| S-EPMC11004130 | biostudies-literature
| S-EPMC11799470 | biostudies-literature