Impact of total variation regularized expectation maximization reconstruction on the image quality of 68Ga-PSMA PET: a phantom and patient study.


ABSTRACT:

Objectives

To investigate the impact of total variation regularized expectation maximization (TVREM) reconstruction on the image quality of 68Ga-PSMA-11 PET/CT using phantom and patient data.

Methods

Images of a phantom with small hot sphere inserts and 20 prostate cancer patients were acquired with a digital PET/CT using list-mode and reconstructed with ordered subset expectation maximization (OSEM) and TVREM with seven penalisation factors between 0.01 and 0.42 for 2 and 3 minutes-per-bed (m/b) acquisition. The contrast recovery (CR) and background variability (BV) of the phantom, image noise of the liver, and SUVmax of the lesions were measured. Qualitative image quality was scored by two radiologists using a 5-point scale (1-poor, 5-excellent).