Utilizing flip angle/TR equivalence to reduce breath hold duration in hyperpolarized 129 Xe 1-point Dixon gas exchange imaging.


ABSTRACT:

Purpose

To reduce scan duration in hyperpolarized 129 Xe 1-point Dixon gas exchange imaging by utilizing flip angle (FA)/TR equivalence.

Methods

Images were acquired in 12 subjects (n = 3 radiation therapy, n = 1 unexplained dyspnea, n = 8 healthy) using both standard (TR = 15 ms, FA = 20°, duration = 15 s, 998 projections) and "fast" (TR = 5.4 ms, FA = 12°, duration = 11.3 s, 2100 projections) acquisition parameters. For the fast acquisition, 3 image sets were reconstructed using subsets of 1900, 1500, and 1000 projections. From the resulting ventilation, tissue ("barrier"), and red blood cell (RBC) images, image metrics and biomarkers were compared to assess agreement between methods.

Results

Images acquired using both FA/TR settings had similar qualitative appearance. There were no significant differences in SNR, image mean, or image SD between images. Moreover, the percentage of the lungs in "defect", "normal", and "high" bins for each image (ventilation, RBC, barrier) was not significantly different among the acquisition types. After registration, comparison of 3D image metrics (Dice, volume similarity, average distance) agreed well between bins. Images using 1000 projections for reconstruction had no significant differences from images using all projections.

Conclusion

Using flip angle/TR equivalence, hyperpolarized 129 Xe gas exchange images can be acquired via the 1-point Dixon technique in as little as 6 s, compared to ~15 s for previously reported parameter settings. The resulting images from this accelerated scan have no significant differences from the standard method in qualitative appearance or quantitative metrics.

SUBMITTER: Niedbalski PJ 

PROVIDER: S-EPMC8776583 | biostudies-literature | 2022 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Utilizing flip angle/TR equivalence to reduce breath hold duration in hyperpolarized <sup>129</sup> Xe 1-point Dixon gas exchange imaging.

Niedbalski Peter J PJ   Lu Junlan J   Hall Chase S CS   Castro Mario M   Mugler John P JP   Shim Yun M YM   Driehuys Bastiaan B  

Magnetic resonance in medicine 20211013 3


<h4>Purpose</h4>To reduce scan duration in hyperpolarized <sup>129</sup> Xe 1-point Dixon gas exchange imaging by utilizing flip angle (FA)/TR equivalence.<h4>Methods</h4>Images were acquired in 12 subjects (n = 3 radiation therapy, n = 1 unexplained dyspnea, n = 8 healthy) using both standard (TR = 15 ms, FA = 20°, duration = 15 s, 998 projections) and "fast" (TR = 5.4 ms, FA = 12°, duration = 11.3 s, 2100 projections) acquisition parameters. For the fast acquisition, 3 image sets were reconstr  ...[more]

Similar Datasets

| S-EPMC11782336 | biostudies-literature
| S-EPMC5836876 | biostudies-literature
| S-EPMC9919357 | biostudies-literature
| S-EPMC6750965 | biostudies-literature
| S-EPMC10206636 | biostudies-literature
| S-EPMC10629838 | biostudies-literature
| S-EPMC5339783 | biostudies-literature
| S-EPMC6291357 | biostudies-literature
| S-EPMC8178811 | biostudies-literature
| S-EPMC11680735 | biostudies-literature